HI-99301 EC/TDS/Temperature Meter - High Range
HI-99301 EC/TDS/Temperature Meter - High Range
Price excludes VAT.
Hurry up, only 3 items left in stock.
Couldn't load pickup availability
Description
Description
The HI-99301 is a portable, EC/TDS/temperature meter that measures high range conductivity in mS/cm and TDS in ppt. Both instruments are housed in a case rated for IP67 conditions.
This instrument offers single point, automatic calibration with automatically temperature compensated measurements. The compensation coefficient and EC/TDS conversion factor are user-selectable from 0.45 to 1.00 and temperature coefficient is selectable from 0.0 to 2.4%/°C.
This meter easily fits in the palm of your hand and the bottom probe connection ensures the electrode cable doesn't get in your way. The large, multi-level LCD displays the primary reading, temperature and calibration guides simultaneously. Symbols and messages on the LCD indicate meter status and guides users through operations.
At start-up, the meter shows the remaining battery percentage and when a low battery condition is detected, a battery symbol appears on the LCD to advise the user that only a few hours of working time is left.
The HI-76306 EC/TDS/temperature probe resists clogging and is easy to clean.
Ideal for: Boiler Water & Waste Water
Main features
- Simultaneous EC/TDS and temperature measurements on a large LCD display
- Selectable temperature unit
- HI-763063 dedicated EC/TDS temperature probe
- Probe quick connect system
- Battery life indication and low battery detection
- Auto-off function
- Keystroke confirmation tone
- Waterproof casing IP67
Each meter is supplied complete with:
- HI-763063 pH/EC/TDS probe with built-in temperature sensor, DIN connector and 1m cable
- HI-70030 – 12880 µS/cm and HI-70038 6.44 ppt 1 sachet each
- 100mL beaker
- Alkaline batteries: 1.5V AAA x 3
- Rugged carrying case
- Calibration certificate of meter
- Calibration certificate of probe
- Instruction manual
Specifications
Specifications
| Range | ||
| EC | 0 to 20.00 mS/cm (displays mS) | |
| TDS | 0.00 to 10.00ppt | |
| Temp | -5.0 to 105.0°C | |
| Resolution | ||
| EC | 0.01 mS/cm | |
| TDS | 0.01 ppt (g/L) | |
| Temp | 0.1°C | |
| Accuracy @ 20°C | ±2% F.S. (EC/TDS); ±0.5°C temp | |
| Temperature Compensation | Automatic, with β selectable from 0.0 to 2.4%/°C with 0.1 increments | |
| EC/TDS Calibration | Automatic, one-point at 12.88mS/cm or 6.44 ppt (CONV=0.5) or 9.02 ppt (CONV=0.7) | |
| TDS Conversion factor | Selectable from 0.45 to 1.00 with 0.01 increments | |
| Probe (included) | HI-763063 EC/TDS/Temperature sensor, DIN connector and 1m cable | |
| Battery type/life | 1.5V AAA x 3 approx 500 hours of continuous use | |
| Auto off | User selectable after 8 mins, 60 mins or disabled | |
| Environment | 0 to 50°C RH max 100% | |
| Meter Dimensions | 154 x 63 x 30mm | |
| Weight (with batteries) | 196g | |
| Casing ingress protection rating | IP67 |