HI-98294 pH/EC/ISE/Turbidity/DO Meter + GPS + 4m cable
HI-98294 pH/EC/ISE/Turbidity/DO Meter + GPS + 4m cable
Hurry up, only 2 items left in stock.
The Hanna HI-98294 is an advanced GPS-enabled multiparameter water quality meter designed for environmental consultants, water authorities, laboratories, water treatment facilities and aquaculture professionals.
Ideal for monitoring rivers, lakes, groundwater, coastal waters, reservoirs and aquaculture environments, the HI-98294 measures a comprehensive range of water quality parameters, including pH, ORP, conductivity, TDS, resistivity, salinity, dissolved oxygen, turbidity and temperature. Additional ion-selective measurements are available using optional sensors.
Its integrated 12-channel GPS receiver records geographical coordinates alongside measurement data, making it easy to identify sampling locations and compare results over time. The field-replaceable sensors feature automatic recognition, simplifying maintenance and setup.
Designed for demanding field conditions, the meter has an IP67-rated housing and connects to an IP68-rated probe capable of operating at depths of up to 20 metres. Data logging stores up to 50,000 interval measurements, while USB-C connectivity allows convenient data transfer for further analysis.
Couldn't load pickup availability
Specifications
Specifications
Full Description Specification
pH / mV
| Range | 0.00 to 14.00 pH / ±600.0 mV |
| Resolution | 0.01 pH / 0.1 mV |
| Accuracy | ±0.05 pH / ±3.0 mV |
| Calibration | 1-point quick calibration or up to 3 points (pH 4.01, 6.86, 7.01, 9.18, 10.01) or custom |
ORP
| Range | ±2000.0 mV |
| Resolution | 0.1 mV |
| Accuracy | ±10 mV |
| Calibration | Automatic, 1 custom point |
EC (Conductivity)
| Range | 0–200 mS/cm (400 mS/cm absolute) |
| Resolution | Auto ranging from 1 μS/cm to 0.1 mS/cm |
| Accuracy | ±1% of reading or ±1 μS/cm |
| Calibration | 1-point quick or standards: 84 μS, 1413 μS, 5.00, 12.88, 80.0, 111.8 mS/cm |
TDS
| Range | 0 to 400,000 ppm |
| Resolution | Auto ranging ppm / ppt |
| Accuracy | ±1% of reading or ±1 ppm |
| Calibration | Based on EC or salinity |
Resistivity
| Range | Up to 1.0000 MΩ·cm |
| Resolution | 1 Ω·cm to 0.0001 MΩ·cm |
| Calibration | Based on EC or salinity |
Salinity
| Range | 0.00 to 70.00 PSU |
| Resolution | 0.01 PSU |
| Accuracy | ±2% or ±0.01 PSU |
| Calibration | 1-point custom |
Turbidity
| Range | 0.0–1000 FNU |
| Resolution | 0.1 FNU / 1 FNU |
| Accuracy | ±0.3 FNU or ±2% |
| Calibration | Up to 3 points (0, 20, 200 FNU) |
Dissolved Oxygen
| Range | 0–500% / 0.00–50.00 mg/L |
| Resolution | 0.1% / 0.01 mg/L |
| Accuracy | ±1.5% (0–200%), ±5% (higher range) |
| Calibration | Air calibration, 1–2 points or custom |
| Pressure Compensation | Automatic |
Temperature
| Range | -5.00 to 50.00°C |
| Resolution | 0.01°C |
| Accuracy | ±0.15°C |
| Calibration | 1-point |
Additional Specifications
| Logging | 50,000 records (interval) / 15,000 on-demand |
| Logging Interval | 1 sec to 3 hours |
| GPS | Yes, 12 channel |
| Connectivity | USB-C (host & device) |
| Protection | IP67 |
| Battery | 4x AA, ~40 hours |
| Dimensions | 185 x 93 x 35.2 mm |
| Weight | 435 g |
Probe – HI-7698294
| Inputs | pH, ORP, DO, EC, Turbidity |
| Environment | Fresh, brackish, seawater |
| Protection | IP68 |
| Max Depth | 20 m |
| Temperature | -5 to 50°C |
Sensors
| pH/ORP | 0–12 pH, ±2000 mV, glass + platinum |
| EC/Turbidity | Up to 400 mS/cm, 1000 FNU |
| DO | 0–500% / 0–50 mg/L |
Safety Data Sheet
Safety Data Sheet
Manuals
Manuals
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.